Carlos Ojeda, Danicsa Espino, Diego Henrri Ojeda Alberca, Zayra Marivel More Davis, Fátima Giovana Vásquez Vigil, Fadia Zulema Huerta Huerta, Flor Juliana Mercedes Risco Ipanaque, Samame Talledo Belinda Elvira (2024). Precision and reliability of digital blood pressure monitor when monitoring the blood pressure of patients in critical units of a hospital in Peru. In: Human Interaction and Emerging Technologies (IHIET-AI 2024): Artificial Intelligence and Future Applications. AHFE Open Access, vol. 120, pp. 491–496. https://doi.org/10.54941/ahfe1004601
The increasing integration of digital technologies in the healthcare field has raised questions about the accuracy of these devices. Therefore, this study aims to support, through scientific evidence, its use in critical environments. The main objective of the article was to evaluate the accuracy and reliability of digital blood pressure monitors used to monitor the blood pressure of patients in critical health units. To carry out the analysis, a qualitative approach was adopted that involved the review and analysis of a set of ten previously published studies, which were selected from highly reliable academic and medical sources. The choice of a qualitative approach was based on the need to carry out a thorough and detailed examination of the studies available in the scientific literature, with the purpose of identifying patterns, trends and limitations in the use of digital blood pressure monitors in critical healthcare settings.During the review and analysis process of the ten selected studies, special attention was given to several key aspects. Firstly, the methodology used in each of the studies was evaluated, paying particular attention to their methodological quality. Likewise, the measurement techniques used, the sample size, the population studied and the conditions were examined. environments in which the measurements were carried out. The results obtained from this analysis revealed interesting trends regarding the accuracy and reliability of digital blood pressure monitors in blood pressure monitoring. Overall, digital devices demonstrated satisfactory accuracy and reliability in blood pressure measurement. However, it is important to highlight that some limitations and variations in the results are identified, which can be attributed to various factors. One of the factors identified as influencing the precision of the measurements was the obesity of the patients, since in these cases digital blood pressure monitors can provide less precise measurements. Additionally, the position of the arm during measurement can also influence the results, underscoring the importance of standardizing position to obtain consistent and reliable measurements.On the other hand, poor quality cuffs can affect the accuracy of blood pressure measurements. Therefore, the use of high-quality cuffs in clinical practice is strongly recommended. Finally, the environmental conditions under which measurements are made can influence the results. Factors such as temperature and humidity can affect the function of digital blood pressure monitors, highlighting the importance of controlling these conditions for accurate and reliable measurements. In conclusion, digital blood pressure monitors have emerged as valuable tools in monitoring blood pressure in critical health units. However, its successful implementation requires careful evaluation of the circumstances and precise calibration.
The growing integration of digital technologies in the field of health has raised doubts about the accuracy of devices that measure blood pressure dig- itally, these are used to monitor blood pressure, they consist of a band that fits on the arm (cuff) and a device that records pressure (Arana C., 2022). Taking into account the aforementioned, it is evident that technological advances have led to various work processes in the healthcare environment
being carried out much more quickly, which is why the Pan American Health Organization (PAHO) considers that this new technology is a support in optimizing time in health care and in monitoring blood pressure by health personnel. This technological adaptation implies feasibility, but it has also generated a new learning challenge for some health professionals, who see this digitalization as a bit complicated, either because of the year in which they learned to measure blood pressure or because of the simple fact. of mech- anized averse with the use of the manual tensiometer (Lizcano-Jaramillo, P. A., 2019).
On the other hand, there is evidence that the common use of automatic devices to measure blood pressure in critical health settings has raised con- cerns about their comparability compared to manual measurements with mercury sphygmomanometers, traditionally known as standard (Ostchega, et al., 2012). This is why Borneo (2020), in its study on the self-perception of digital skills in the Health personnel of Huánuco in Peru, found that most public workers had a basic level of digital skills, in addition, they excelled in the dimension technological, followed by digital participation, along with collaboration and communication, on the one hand, by mastering the four necessary aspects, only a small group felt digitally competent.
Additionally, it is understood that there are factors that affect the accuracy and reliability of these devices in a critical clinical environment, including the type of technologies used, the training of medical personnel, and the competence in the use of these devices to measure blood pressure in units reviews (MediPlus, 2023). Therefore, Padwal et al. (2020) points out that there are numerous problems, including the skills and abilities of health per- sonnel working in these units, as they may be possible obstacles to ensuring accurate blood pressure measurement with monitors, this emphasizes that there is still health personnel who do not adapt to technological devices. However, the accuracy and reliability of digital blood pressure monitors in monitoring blood pressure in critical units is an essential aspect of mod- ern healthcare that requires careful attention and a greater understanding of the competencies and technologies involved, thus leading to techno- logical advance to a point of help for improving health (Cardona et al., 2018).
METHODOLOGICAL FRAMEWORK Type of Investigation
The present literature review study was descriptive, because it is the most appropriate way to gather data that evidence the connections of the world in its natural state. This type of analysis is carried out as a first step before undertaking an experiment, allowing the precise identification of which variables influence and which elements they incorporate in the experimen- tal study (Muguira, A., 2018). That is why the study has a qualitative approach, since the bibliographic works found are mostly qualitative, but we also found quantitative studies, so we proceed with descriptive and numerical data.
Precision and Reliability of Digital Blood Pressure Monitor 493
Methodology
A qualitative approach was adopted that involved the review and analysis of ten previously published studies from reliable sources (PUBMED, SCIELO, RECIMUNDO, REPOSITORIO.WIENER). In addition, an advanced inves- tigation was carried out using various search tactics, where descriptors and keywords were combined. Considering language and age as filters (English and Spanish, 15 years, respectively). The sample was made up of 10 articles, of which 7 belong to qualitative studies and 3 to quantitative studies.
RESULTS AND DISCUSSION
Of the 100% of the articles selected, 70% were qualitative studies and 30% were quantitative studies (Table 1).
Table 1. Percentage of articles selected according to study methodology.
Table available in the original PDF paper.
Study methodology % Qualitative studies 70% Quantitative studies 30% Total 100%
After a thorough analysis of the 10 original articles, the identified themes were classified into three different categories using a thematic analysis approach. These categories were named as follows: The first category refers to the comparison of digital devices in PA taking, compared to those that traditionally exist. For this context Ostchega et al. (2012) indicates that the differences between the devices (automatic oscillometric and mercury sphyg- momanometers) were minimal or close to 2 mm Hg in all subgroups, whether by gender, age group, ethnicity, body mass index categories and variations in cuff size. However, exceptions were seen in the mean differences in systolic blood pressure, particularly in those using the extra-large blood pressure cuff (−3.1 mm Hg) and in individuals with obesity (−2.6 mm Hg), as well as in blood pressure.. diastolic blood pressure for the low weight group (−3.5 mm Hg). While Araujo et al. (2022) mentions that systolic BP measured by automatic blood pressure monitors is 1.17 mmHg on average in 95%; for diastolic BP, it produced −0.08 mmHg the average is 95% ranging between 0.69 and 0.54, compared to a mercury sphygmomanometer, so automatic blood pressure monitors have strong validity measurement compared to the column of mercury. Therefore, they can be safely used to measure blood pressure in children and adolescents in clinical and epidemiological studies. These findings, when compared with the study by Park, S.-H., & Park, Y.-S. (2019), found that automatic oscillometric devices recorded lower readings compared to mercury sphygmomanometers for both systolic blood pressure, with a mean difference of 1.75 mmHg (95% confidence interval: −3.05 to −0.45, I = 91.0%), as for diastolic blood pressure, with a mean difference of −1.20 mmHg (95% confidence interval: −2.16 to −0.24, I = 95.0%). When analyzed by manufacturer, the BpTRU device showed lower measurements
compared to the mercury sphygmomanometer, while the OMRON showed no significant differences in systolic and diastolic blood pressure measure- ments compared to the mercury sphygmomanometer, although the Results will vary depending on device used.
The second category refers to the accuracy of digital monitors in measur- ing the blood pressure of critically ill patients. Corresponding to this, Kamboj et al. (2021) states that the intervals of agreement in the population for sys- tolic blood pressure were widely extended, varying from −36.13 mmHg to 28.28, in terms of the precision of diastolic blood pressure measurements, they showed a notable lack of consistency, which which resulted in unreli- able estimates of population concordance intervals. Regarding non-invasive continuous blood pressure measurement based on tonometry, Saugel et al. (2013) point out that they examined and contrasted blood pressure values in a total of 4,502 records, each representing the average of 10 heartbeats. A positive bias of +0.72 mm Hg was detected in the Mean Blood Pressure (MAP), with a range of 95% agreement that varied between −9.37 mm Hg and +10.82 mm Hg, where mean differences were observed of −1.39 mm Hg and +4.36 mm Hg, respectively. Regarding automatic devices for selfmea- surement of blood pressure according to the International Protocol Asmar et al. (2010), tested three devices, successfully passing both the first and second stages of the validation process. The average discrepancies between the devices and mercury measurements were around 1.9±3.0 and −1.0±2.3 mmHg for systolic and diastolic blood pressure, respectively, for the Omron M3 device. Intellisense; 2.5±5.4 and −2.3±3.6 mmHg for the Omron M2 Compact device; and 1.4±4.5 and 0.8±4.6 mmHg for the Omron R3-I Plus device, thus, meeting the requirements of the International Protocol, they can be used to measure blood pressure.
The third category refers to the use of the digital monitor to take blood pressure. Regarding Asian hypertension management: current status, home blood pressure, and specific concerns in Indonesia, Turana et al. (2020) inves- tigated that home blood pressure monitoring, known as LMWH, where they mention that it is a valuable tool since 70% of the diagnosis in the control of hypertension is more effective. However, in Indonesia, access to LMWH is restricted by factors such as the cost of the devices and a widespread percep- tion that these devices lack reliability. Finally, with regard to the reliability of automated devices, Cardona et al. (2018) conducted a study in which 56 patients over 19 years of age were included, of which nineteen (38%) had high blood pressure. All of these patients had their blood pressure mea- sured using both a traditional and a digital device. Aiming to evaluate the reliability of the automated device. The results indicated that the automated device proved to be a reliable tool for measuring blood pressure.
CONCLUSION
In the process of reliability of automated devices, studies compared blood pressure measurements obtained with an automated device and a mercury sphygmomanometer where it was found that there were no significant differ- ences in systolic or diastolic blood pressure measurements between the two
Precision and Reliability of Digital Blood Pressure Monitor 495
methods. This suggests that automated devices, such as the OMRON HEM- 7320-LA, can provide reliable measurements comparable to those obtained with traditional methods.
On the other hand, the articles read state that the use of automatic devices in critical emergency patients has a similar effectiveness to invasive mea- surement, which suggests that these devices can be a valid and less invasive alternative in emergency medical care settings.
Regarding the precision in measuring children and adolescents, the research revealed that automatic devices for measuring blood pressure are accurate and valid in different clinical and epidemiological studies. This is important as these devices can provide reliable measurements in a more efficient manner.
Differences between automatic devices and mercury sphygmomanometers are also presented, where several studies showed that blood pressure mea- surements between these devices, especially in certain age groups and with serious medical conditions, highlight the importance of considering the type of device used when interpreting measurements and control of hypertension.
Finally, this non-invasive monitoring in critical patients demonstrates how important it is to be constantly observed when measuring blood pressure in critical patients. Taking into account that this is a much more accessible alter- native, thus avoiding invasive procedures, this is why advances in technology allow valuable information to be obtained in a less invasive way, which can improve patient care and optimization, critics.
RECOGNITION Thanks to the participants (co-authors) for their support in the thorough search of high-impact articles for the development of the research.
REFERENCES
Arana, C. (2022, February 7). Digital blood pressure monitor: what it is, what it is used for and why you need it. Editions EL PAÍS S. L. https://elpais.com/ mexico/escaparate/estilo-de-vida/2022-02-07/tensiometrodigital-que-es-para- que-se-usa-y-por-que-lo-necesitas.html Araujo-Moura, K., Souza, L. G., Mello, G. L., & De Moraes, A. C. F. (2022).
Blood pressure measurement in pediatric population: comparison between automated oscillometric devices and mercury sphygmomanometers—a system- atic review and meta-analysis. European Journal of Pediatrics, 181(1), 9–22. https://doi.org/10.1007/s00431-021-04171-3 Asmar, R., Khabouth, J., Topouchian, J., El Feghali, R., & Mattar, J. (2010). Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intel- lisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Pressure Monitoring, 15(1), 49–54. https://doi.org/10.1097/mbp.0b013e3283354b11 Borneo E. (2020). Self-perception of digital competencies in administrative staff of the Huánuco Health Network. Scientific Journal of Social Sciences and Humanities, 11(1), 19–24. https://bit.ly/3Zzt0jU
Cardona-Müller, D., Grover-Páez, F., Guzmán-Saldivar, V., Alanis-Sánchez, G. A., Murguía-Soto, C., Totsuka-Sutto, S. E., Quezada-Fernández, P., MacíasChumac- era, A., Cardona-Muñoz, E. G., & Ramos-Becerra, C. G. (2018). Reliability of an automated monitor for blood pressure measurement. Medical Journal of Chile, 146(2), 190–195. https://doi.org/10.4067/s003498872018000200190 Dialnet-InvasiveAndNoninvasiveMonitoringInPatientsIngres-8056959. Pdf. Retrieved October 5, 2023, from http://Dialnet-MonitorizacionInvasiva YNoInvasivaEnPacientesIngres-8056959.pdf. Kamboj, N., Chang, K., Metcalfe, K., Chu, C. H., & Conway, A. (2021). Accu-
racy and precision of continuous non-invasive arterial pressure monitoring in critical care: A systematic review and meta-analysis. Intensive & Critical Care Nursing: The Official Journal of the British Association of Critical Care Nurses, 67(103091), 103091. https://doi.org/10.1016/j.iccn.2021.103091 Lizcano-Jaramillo, P. A. (2021). Health Technology Assessment: A Hos- pital Approach to the Incorporation of Medical Devices. Mexican Journal of Biomedical Engineering. https://www.scielo.org.mx/scielo.php? script=sci_arttext&pid=S0188-95322019000300010 Muguira, A. (2018, October 23). What is descriptive research? QuestionPro.
https://www.questionpro.com/blog/es/investigacion-descriptiva/ Ostchega, Y., Zhang, G., Sorlie, P., Hughes, J. P., Reed-Gillette, D. S., Nwankwo, T., & Yoon, S. (2012). Blood pressure randomized methodology study compar- ing automatic oscillometric and mercury sphygmomanometer devices: National Health and Nutrition Examination Survey, 2009–2010. National Health Statistics Reports, 59. https://pubmed.ncbi.nlm.nih.gov/24984529/ Padwal, R., Campbell, N. R. C., Schutte, A. E., Olsen, M. H., Delles, C., Etyang, A.,
Cruickshank, J. K., Stergiou, G., Rakotz, M. K., Wozniak, G., Jaffe, M. G., Ben- jamin, I., Parati, G., & Sharman, J. E. (2020). Optimizing observer performance when measuring office blood pressure: position statement of the Lancet Com- mission on Hypertension. Revista panamericana de salud publica [Pan American journal of public health], 44, 1. https://doi.org/10.26633/rpsp.2020.88 Park, S.-H., & Park, Y.-S. (2019). Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a system- atic review and meta-analysis. Blood Pressure Monitoring, 24(6), 265–276. https://doi.org/10.1097/mbp.0000000000000412 Sagirova, Z., Kuznetsova, N., Gogiberidze, N., Gognieva, D., Suvorov, A.,
Chomakhidze, P., Omboni, S., Saner, H., & Kopylov, P. (2021). Cuffless blood pressure measurement using a smartphone-case based ECG monitor with photo- plethysmography in hypertensive patients. Sensors (Basel, Switzerland), 21(10), 3525. https://doi.org/10.3390/s21103525 Saugel, B., Meidert, A. S., Hapfelmeier, A., Eyer, F., Schmid, R. M., & Huber, W.
(2013). Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. British Journal of Anaesthesia, 111(2), 185–190. https://doi.org/10.1093/bja/aet025 Turana, Y., Tengkawan, J., & Soenarta, A. A. (2020). Asian management of hypertension: Current status, home blood pressure, and specific concerns in Indonesia. Journal of Clinical Hypertension (Greenwich, Conn.), 22(3), 483–485. https://doi.org/10.1111/jch.13681
LISTED IN THE AHFE DIGITAL BOOKSTORE CATALOGUE
Explore the complete digital volume
Human Interaction and Emerging Technologies (IHIET-AI 2024): Artificial Intelligence and Future Applications
Read this individual research paper free through the original publisher, or explore the full digital proceedings book.
ISBN 978-1-958651-96-4
Explore this ISBN in the AHFE Bookstore ↗Listed in the July 2026 bookstore catalogue; search by ISBN to confirm current availability · Digital book purchase optional