Claus Backhaus, Simon Siebers (2022). Work-related noise exposure in a neonatal intensive care unit. In: Healthcare and Medical Devices. AHFE Open Access, vol. 51, pp. 286–292. https://doi.org/10.54941/ahfe1002129
Premature infants are children born before the 37th week of gestation. They often need to be cared for in neonatal intensive care units (NICUs) after birth, where they are continuously exposed to noise that can affect sleep patterns and lead to growth and developmental delays. To date, there are no unified limits for the prevention of noise exposure in NICUs. Recommendations for maximal continuous sound levels vary - depending on the professional society - between 35 and 60 dB(A) for daytime values and 20 to 60 dB(A) for nighttime values. They span a range from quiet whispering to normal road traffic noise. Existing studies of noise measurements in NICUs indicate that these recommendations are generally not met. A limitation of previous work is their short time periods that noise measurements were made for, usually one or two hours. As a result, it is not possible to make statements about the diurnal cyclic change in noise exposure, which may have an influence on the circadian rhythm of premature infants, for example. The present work aims to determine the intensity and diurnal cyclic differences of noise exposure in a NICU. In addition, particularly noisy work activities are identified in order to derive suggestions for prevention.For this purpose, the continuous sound levels are recorded for 22 work shifts in a German level 1 perinatal center. Measurements are made in the patient room and the inside of an infant incubator. The mean daily noise exposure levels (LEX,8h) of the early, late and night shifts are calculated and the peak sound level (LpCpeak) is recorded. To test for a day-night rhythm, these are compared using one-factor ANOVA. Using the Contextual Inquiry method, work analyses are performed for 12 work shifts and the early, late and night shifts with the highest equivalent continuous sound level (LAeq) are examined for noise-intensive work activities as examples. The measurement results in LEX,8h and LpCpeak values of up to 49.7 dB(A) and 124.4 dB(C) inside the infant incubator and 55.4 dB(A) and 110.3 dB(C) in the patient room. A day-night rhythm of noise exposure could not be demonstrated. The most noise-intensive work activities included the suctioning of patients and the conversations conducted in the process, with an LAeq of 63 dB(A). In addition, numerous very high impulsive noise events are identified which contribute to the noise exposure of premature infants. The presented study proves that the noise exposure is too high, which is very likely to affect the sleep behavior of premature infants. The results correspond with the statements of comparable studies. It is remarkable that especially the high peak noise levels are mainly due to trivial causes, which can be avoided by simple preventive measures.
Preterm infants are children born before the 37th gestation week. After birth, they often have to be cared for in neonatal intensive care units (NICUs), where they are continuously exposed to noise (Boggini et al. 2020). Prema- ture infants are in a particularly vulnerable exposure group, as noise impairs sleep patterns and leads to growth and developmental delays (Williams et al. 2009). Noise can also have detrimental effects on the cardiovascular and respiratory systems (Boggini et al. 2020, Romeu et al. 2016, Schefer Cardoso et al. 2015, Wachmann & Lahav 2011) and lead to psychological distress in preterm infants, manifested, for example, in delayed language development (Marik et al. 2012, McMahon et al. 2012, Stromswold & Sheffield 2004). Because of these extensive health effects of noise, the American Academy of Pediatrics (1997) recommends that continuous sound levels should not exceed 45 dB(A) in NICUs. However, noise measurements in NICUs show that this level is often not met. Darcy et al. (2008), Krueger et al. (2005), Parra et al. (2017), and Lasky and Williams (2009) measured sound levels
Work-Related Noise Exposure in a Neonatal Intensive Care Unit 287
averaging 54 to 68 dB(A). Only a study by Domanico et al. (2011) measured an equivalent continuous sound level (Leq) of 33 dB(A) in a single room after remodeling a NICU.
A limitation of the previous studies is that the noise measurements were not carried out for an entire work shift. This makes it impossible to make statements about the diurnal rhythm of noise exposure, which can have an influence on the circadian rhythm of premature infants, for example. Since noise affects the sleep behavior of preterm infants (Kuhn et al. 2013) and a disturbed sleep-wake cycle has effects on thermoregulation, the immune system, and hormone production (Schultes & Fehm 2004), an insufficient day-night rhythm can impair the developmental process (Schrader & Schra- der 2001). In addition, noise affects heart rate (Long et al. 1980, Vranekovic et al. 1975), blood pressure (Jurkovicová & Aghová 1989, Williams et al. 2009), oxygen saturation (Johnson 2001, Zahr & Balian 1995), and respira- tory rate (Wharrad & Davis 1997) of preterm infants. However, the intensity and direction of these effects are not homogenous in previous studies, which is due to the fact that mostly the physiological response to high noise stimuli of 80 to 100 dB(A) has been studied (Field et al. 1979, Jurkovicová & Aghová 1989, Vranekovic et al. 1974, Wharrad & Davis 1997). Studies on natural noise exposure in NICUs and its effect on preterm infants are not sufficiently available so far (Wachman & Lahav 2011).
The present work aims to determine the intensity and diurnal rhythm differences of noise exposure in a German NICU.
METHOD
To record the noise exposure, the sound levels are measured for 22 work shifts in the patient room of a German perinatal center (level 1). The selected patient room has four treatment places, all of which are occupied at the time of the measurement. One premature baby is in an incubator and is being ven- tilated non-invasively (CPAP ventilation). Three premature babies are being cared for in warming or crib beds. The measurements are performed accor- ding to ISO 9612 (2009). Two class 1 sound level meters according to IEC 61672-1 (2014) from NTi Audio (Liechtenstein) are used (sound level meter: XL2; microphone: MA2230; sound calibrator: CAL200). The A-weighted equivalent continuous sound level (LAeq) is recorded as a time average of the sound level over the measurement time and the C-weighted peak sound level (LpCpeak), each with a sampling frequency of 10 Hz. The measuring instrument is calibrated before and after each measurement. The measure- ments are performed simultaneously with two identical sound level meters in the incubator and in the patient room. In order not to interfere with the workflow in the patient room, one microphone is fixed to a wall (height 160 cm). The second microphone is covered with a hygienic protective cover for ultrasound probes (20 × 200 mm, Dahlhausen company), attached to the outer wall of the incubator with a suction cup stand and positioned with the microphone tip near the head of the premature baby (distance approx. 20 cm).
288 Backhaus and Siebers
To evaluate the measurements, the daily noise exposure levels (LEX,8h) are calculated from the LAeq. They correspond to the average sound level of an eight-hour work shift and allow direct comparison and further pro- cessing of the results. The arithmetic mean and standard deviation (aM ± SD) of LEX,8h are determined for the early, late, and night shifts, and an independent-samples t-test (α = .05) is used to test whether the incubator and patient room LEX,8h differ. If there is a significant difference, Cohen’s effect size d is determined. A one-sample t-test (α = .05) checks for compli- ance with the American Academy of Pediatrics (1997) recommendations. To test for a day-night rhythm of noise exposure, the association between the averaged LEX,8h of the early, late, and night shifts in the patient room and incubator, respectively, is determined using one-sample ANOVA (α = .05). All statistical data analyses were performed using SPSS Statistics 24 software (IBM Corp., USA).
RESULTS
The sound levels of 8 early, 7 late and 7 night shifts were recorded with a total measurement time of 176 hours. Table 1 shows the calculated LEX,8h (aM ± SD) and LpCpeak (aM ± SD) in the patient room and incubator. The highest measured LpCpeak was 124.4 dB(C) in the incubator and 110.3 dB(C) in the patient room.
The recorded sound levels were normally distributed (Shapiro-Wilk tests, p > .05). The one-sample t-tests performed demonstrate a highly significant difference between the recorded LEX,8h in the patient room and the recom- mended guideline value of 45 dB(A) (t = 23.628, p < .001, n = 24, aM = 54.3, SD = 1.9). The same is true for LEX,8h in the incubator (t = 7.583, p < .001, n = 24, aM = 48.8, SD = 2.4).
The performed single factorial ANOVA confirms that the averaged LEX,8h values in the patient room do not differ significantly between work shifts (F(2, 21) = 2.821, p = .082, partial η2 = .212, n = 24). The same is true for the averaged LEX,8h values in the incubator (F(2, 21) = 1.373, p = .275, partial η2 = .116, n = 24). Thus, a day-night rhythm of noise exposure could not be detected in the studied NICU.
The LEX,8h averaged over all work shifts was 54.3 ± 1.9 dB(A) for the patient room and 49.5 ± 2.4 dB(A) for the incubator. The independent
Table 1. Result of sound level measurement of 22 working shifts of a NICU.
Table available in the original PDF paper.
Sound level Work shift n Incubator* Patient room*
LEX,8h [dB(A)]
early 8 49.7 ± 2.8 55.4 ± 2.0 late 7 47.8 ± 2.2 53.9 ± 1.6 night 7 48.7 ± 1.9 53.3 ± 1.6 LpCpeak [dB(C)] early 8 113.6 ± 7.7 104.2 ± 3.8 late 7 107.6 ± 4.4 104.1 ± 4.1 night 7 107.4 ± 9.2 101.6 ± 1.6
*aM ± SD
Work-Related Noise Exposure in a Neonatal Intensive Care Unit 289
samples t-test performed shows that the averaged LEX,8h is significantly higher in the patient room (aM = 54.3, SD = 1.9, n = 24) than in the incuba- tor (aM = 49.5, SD = 2.4, n = 24), t(46) = −8.156, p < 0.001. The Cohen’s effect size is d = .77, corresponding to a strong effect.
CONCLUSION
The results of the sound level measurements show that all measured LEX,8h exceed the recommendation of the American Academy of Pediatrics (1997), by at least 3 dB(A). Premature infants are exposed to excessive noise levels in both the incubator and patient room (e.g., warming bed or crib). In the full-shift measurements the results of previous studies are thus confirmed, in which - with shorter measurement times - the recommended sound level of 45 dB(A) was also exceeded (cf. Darcy et al. 2008, Krueger et al. 2005, Lasky & Williams 2009, Parra et al. 2017). A common cause of noise is conver- sations that occur between employees (Baker 1992, Connor & Ortiz 2009, Gabor et al. 2003, Morrison et al. 2003, Overman et al. 2008). When these occur during noisy work activities, such as suctioning of preterm infants, they tend to be conducted at a slightly higher pitch and louder. This phenomenon, known as the Lombard effect (Klatte et al. 2010), leads to a self-amplification of the noise exposure. To avoid this, conversations should be avoided as far as possible in NICUs during noise-intensive work activities. Another cause of noise are alarms from medical devices, which were also identified in pre- vious studies by Carvalho et al. (2005), Morrison et al. (2003) and Gabor et al. (2003). Because of the high number of false alarms, nurses find them particularly disturbing (Lawless 1994). A simple possibility for prevention would be the use of radio message receivers (pagers), which would only tran- smit optical, instead of optical-acoustic device alarms, to the staff, which would further reduce the noise in the patient room (cf. Imhoff et al. 2009). As expected, LEX,8h are lower in the incubator than in the patient room. The difference is on average about 6 dB(A). Striking are the high LpCpeak values in the incubator, which with the highest measured peak sound level of 124.4 dB(C) are already close to the pain threshold for adults and are not rea- sonable for premature infants. The measured high LpCpeak values are caused by placing work equipment on the incubator while feeding or caring for the premature infants. The incubator can act as a resonance body and amplify the sound impulse introduced. To reduce these high peak sound levels, the placing of work equipment on the incubator must be avoided as a matter of urgency. In order to achieve this, there must be sufficient storage space within the nurse’s reach. If this is not the case, it should be provided immediately to prevent noise exposure. This problem should also be given special conside- ration in the development of new incubators. For example, the shape of the incubator’s housing dome could prevent work equipment from being placed on it (e.g., by means of a curved surface). An alternative would be to provide a storage facility on the incubator that is acoustically decoupled from the incu- bator door. The evaluation of the sound level measurement shows that there is no significant difference between the LEX,8h of the early, late and night shifts. Thus, a day-night rhythm of noise exposure does not exist either in the
290 Backhaus and Siebers
patient room or in the incubator. It can be assumed that the sleep behavior of premature infants is disturbed by the permanently high noise exposure (Kuhn et al. 2013). To improve sleep quality, rest periods should be introduced in the NICU (Bhutta & Anand 2002), during which lighting levels and nursing activities are also reduced for a certain period of time (Dennis et al. 2010).
Limitations of the presented study include the fact that the sound level measurements and work analyses were only conducted in a German NICU. However, the transferability of the results of the measured LEX,8h values could be proven on the basis of the available literature. Since corresponding publications for LpCpeak values are not available, the transferability of these results should be investigated in further studies. A technical limitation results from the use of the hygienic protective covers for ultrasound probes, which were pulled over the microphone in the incubator. This leads to an attenua- tion of the sound wave hitting the microphone membrane which results in a reduction in the measured sound pressure levels in the incubator. The validity of the measured values in the incubator is thus limited, since higher measu- red values can be expected without the protective covers. However, since the available data already exceed the recommended guideline values of the Ame- rican Academy of Pediatrics (1997), they do not contradict the results of the study.
REFERENCES
American Academy of Pediatrics (1997) Noise: A Hazard for the Fetus and Newborn.
Pediatrics 100:24–727. DOI: 10.1542/peds.100.4.724 Baker CF (1992) Discomfort to environmental noise: heart rate responses of SICU patients. Crit Care Nurs Q. 15(2): 75-90. DOI: 10.1097/00002727-199208000- 00006 Bhutta AT, Anand KJS (2002) Vulnerability of the developing brain. Clin. Perina-
tol. 29(3): 357-372. DOI: 10.1016/S0095-5108(02)00011-8 Boggini T, Pozzoli S, Schiavolin P, Erario R, Mosca F, Brambilla P, Fumagalli M (2020) Cumulative procedural pain and brain development in very preterm infants: A systematic review of clinical and preclinical studies. Neurosci Biobehav Rev 123:320-336. DOI: 10.1016/j.neubiorev.2020.12.016 Carvalho WB, Pedreira ML, de Aguiar MA (2005) Noise level in a pediatric intensive
care unit. J Pediatr 81(6): 495–498. DOI: 10.2223/JPED.1424. Chang YP, Pan YJ, Lin YJ, Chang YZ, Lin CH (2006) A noise-sensor light alarm reduces noise in the newborn intensive care unit. Am J Perinatol 23(5): 265–271. DOI: 10.1055/s-2006-941455. Connor A, Ortiz E (2009) Staff Solutions for Noise Reduction in the Workplace.
Perm J 13(4): 23–27. DOI: 10.7812/tpp/09-057 Darcy AE, Hancock LE, Ware EJ (2008) A descriptive study of noise in the neonatal intensive care unit: ambient levels and perceptions of contributing factors. Adv Neonatal Care 8:16-26. DOI: 10.1097/01.ANC.0000337268.85717.7c. Dennis CM, Lee R, Woodard EK, Szalaj JJ, Walker CA (2010) Benefits of
quiet time for neuro-intensive care patients. J Neurosci Nurs 42(4): 217–224. DOI: 10.1097/JNN.0b013e3181e26c20 Domanico R, Davis DK, Coleman F, Davis BO (2011) Documenting the NICU design dilemma: comparative patient progress in open-ward and single family room units. J Perinatol 31(4): 281–288. DOI: 10.1038/jp.2010.120.
Work-Related Noise Exposure in a Neonatal Intensive Care Unit 291
Field TM, Dempsey JR, Hatsch J, Ting G, Cliffon RK (1979) Cardiac and behavioral responses to repeated tactile and auditory stimulation by preterm and term neonates. Dev Psychol 15(4): 406–416. DOI: 10.1037/0012-1649.15.4.406. Gabor JY, Cooper AB, Crombach SA, Lee B, Kadikar N, Bettger HE, Hanly PJ (2003)
Contribution of the intensive care unit environment to sleep disruption in mecha- nically ventilated patients and healthy subjects. Am J Respir Crit Care Med 167(5): 708–715. DOI: 10.1164/rccm.2201090. IEC 61672-1, 2014–07: Electroacoustics - Sound level meters, Part 1 Specifications.
Berlin: Beuth Verlag ISO 9612, 2009–09: Acoustics - Determination of occupational noise exposure - Engineering method. Berlin: Beuth Verlag Imhoff M, Kuhls S, Gather U, Fried R. (2009) Smart alarms from medical devi-
ces in the OR and ICU. Best Pract Res Clin Anaesthesiol. 23(1): 39–50. DOI: 10.1016/j.bpa.2008.07.008. Johnson AN (2001) Neonatal response to control of noise inside the incubator.
Pediatr Nurs 27(6): 600–605. Jurkovicová J, Aghová L (1989) Evaluation of the effects of noise exposure on various body functions in low-birthweight newborns. Act Nerc Super 31(3): 228–229. Klatte M, Hellbrück J, Seidel J, Leistner P (2010) Effects of Classroom Acoustics
on Performance and Well-Being in Elementary School Children: A Field Study. Environment and Behavior 42(5): 659–692. DOI: 10.1177/0013916509336813. Krueger C, Wall S, Parker L, Nealis R (2005) Elevated sound levels within a busy NICU. Neonatal Netw 24(6): 33–37. DOI: 10.1891/0730-0832.24.6.33. Kuhn P, Zores C, Langlet C, Escande B, Astruc D, Dufour A (2013) Moderate acoustic
changes can disrupt the sleep of very preterm infants in their incubators. Acta paediatrica 102(10): 949–954. DOI: 10.1111/apa.12330. Lasky RE, Williams AL (2009) Noise and light exposures for extremely low birth weight newborns during their stay in the neonatal intensive care unit. Pediatrics 123(2): 540–546. DOI: 10.1542/peds. 2007–3418. Lawless ST (1994) Crying wolf: false alarms in a pediatric intensive care unit. Crit
Care Med 22(6): 981–985. Long JG, Lucey JF, Philip AG (1980) Noise and hypoxemia in the intensive care nursery. Pediatrics 65(1): 143–145. Marik PE, Fuller C, Levitov A, Moll E (2012) Neonatal incubators: a toxic sound
environment for the preterm infant? Pediatr Crit Care Med 13(6): 685–689. DOI: 10.1097/PCC.0b013e31824ea2b7. McMahon E, Wintermark P, Lahav A (2012) Auditory brain development in premature infants. The importance of early experience. Ann N Y Acad Sci 1252:17–24. DOI: 10.1111/j. 1749–6632.2012.06445.x. Morrison WE, Haas EC, Shaffner DH, Garrett ES, Fackler JC (2003) Noise, stress,
and annoyance in a pediatric intensive care unit. Crit Care Med 31(1): 113–119. DOI: 10.1097/00003246-200301000-00018. Overman Dube JA, Barth MM, Cmiel CA, Cutshall SM, Olson SM, Sulla SJ, Nesbitt JC, Sobczak SC, Holland DE (2008) Environmental noise sources and interven- tions to minimize them: a tale of 2 hospitals. J Nurs Care Qual 23(3): 216–224. DOI: 10.1097/01.NCQ.0000324585.68966.51 Parra J, de Suremain A, Berne AF, Ego A, Debillon T (2017) Sound levels in a neona-
tal intensive care unit significantly exceeded recommendations, especially inside incubators. Acta Paediatr 106(12): 1909–1914. DOI: 10.1111/apa.13906.
292 Backhaus and Siebers
Romeu J, Cotrina L, Perapoch J, Linés M (2016) Assessment of environmental noise and its effect on neonates in a Neonatal Intensive Care Unit. Applied Acoustics 111:161–169. DOI: 10.1016/j.apacoust.2016.04.014. Schefer Cardoso SM, de Cássia Kozlowski L, Moreira de Lacerda AB, Marques JM,
Ribas A (2015) Newborn physiological responses to noise in the neonatal unit. Braz J Otorhinolaryngol 81(6): 583–588. DOI: 10.1016/j.bjorl.2014.11.008. Schultes B, Fehm HL (2004) Zirkadiane Rhythmen in der Endokrinologie (circadian rhythms in endocrinology). Internist 45(9): 983-993. DOI: 10.1007/s00108-004- 1249-9 Schrader D, Schrader N (2001) Lärm auf Intensivstationen und dessen Auswirkungen
auf Patienten und Personal (noise in intensive care units and its effects on patients and staff). Intensiv 9(3): 96–106. DOI: 10.1055/s-2001-13064. Stromswold K, Sheffield E (2004) Neonatal Intensive Care Unit Noise & Language Development. Rutgers University Center for Cognitive Science. https://ruccs.rutgers.edu/images/tech_rpt/tr-76/NICU_Noise_TR76.pdf. Access 14. Oktober 2021 Vranekovic G, Hock E, Isaac P, Cordero L (1974) Heart rate variability
and cardiac response to an auditory stimulus. Biol Neonate 24(1): 66–73. DOI: 10.1159/000240633. Wachman EM, Lahav A (2011) The effects of noise on preterm infants in the NICU. Arch Dis Child Fetal Neonatal Ed 96(4): F305-9. DOI: 10.1136/adc.2009.182014. Wharrad HJ, Davis AC (1997) Behavioural and autonomic responses to sound in
preterm and fullterm babies. Br J Audiol 31(5): 315–329. Williams AL, Sanderson M, Lai D, Selwyn BJ, Lasky RE (2009) Intensive care noise and mean arterial blood pressure in extremely low-birth-weight neonates. Am J Perinatol 26(5): 323–329. DOI: 10.1055/s-0028-1104741. Zahr LK, Balian S (1995) Responses of premature infants to routine nursing
interventions and noise in the NICU. Nurs Res 44(3): 179–185.
LISTED IN THE AHFE DIGITAL BOOKSTORE CATALOGUE
Explore the complete digital volume
Healthcare and Medical Devices
Read this individual research paper free through the original publisher, or explore the full digital proceedings book.
ISBN 978-1-958651-27-8
Explore this ISBN in the AHFE Bookstore ↗Listed in the July 2026 bookstore catalogue; search by ISBN to confirm current availability · Digital book purchase optional